Logiclave Clinical Sterilizers
Aimed at the sterilisation of medical devices, the Logiclave Porous Load Sterilizer is intended to sterilize porous and non-porous items by a method using saturated and pressurised steam. It is suitable for steam sterilisation of air entrained porous items such as linen, dressings, wrapped instruments and utensils, hollow instruments and instruments with lumens.Nominal sterilization times and temperatures are:�3 Minutes at 134 - 137DegC�15 Minutes at 121 - 124DegCDesignThe design of the sterilizer is based on the key requirements of the following standards / references:�PD5500�BS EN285�BS 2646�BS 3970 Pt1�BS EN 61010�HTM 2010, HTM2030, CFPP01-01�C14 NHS Model engineering specifications�HSE Guidance Note PM73In addition, the equipment complies fully with the appropriate requirements of the Pressure Equipment Regulations 1999 (SI 1999 No. 2001) and is CE marked accordingly.Vessel constructionThe chamber, jacket and doors are manufactured from minimum 6mm thick high grade 316L stainless steel for maximum corrosion resistance. Other process contact materials are generally of hygienic stainless steel. Each stage of the vessel manufacturing process is witnessed and certified by an independent Insurance Company who also verifies the materials used for construction, welder qualifications, weld procedures and acceptance of hydrostatic pressure test.Control SystemThe sterilizer control system is based on the following hardware components:�Rugged industrial PLC type controller�Analogue input module�Temperature input modules�Digital input modules�Digital output modules�Full colour operator interface (Touch sensitive screen)�Temperature and pressure sensors�Independent data recorder (Where specified)The control system software has been specifically developed for the precision control of the sterilizing process. Program configuration, initiation and monitoring of the process are all possible through the ergonomically designed operator interface.Data RecordingThe Logiclave� Porous Load Sterilizer is supplied with a 40 column thermal printer as standard. Detailed records of each process cycle run including critical times, temperatures and pressures are printed. Program parameters may also be printed for validation records. The Logiclave� system automatically archives to internal memory all logged cycle data for future retrieval, printing and analysis. Typically, up to the last twenty cycles are stored in the database.Further options include independent data recorders with either paper or graphical interfaces provide assurance that sterilizing parameters have been satisfied.Logiclave� SupervisorA PC based data acquisition system that operates over an industry standard Ethernet connection. The Logiclave� Supervisor package can be connected to any number of Logiclave� controllers giving managers the ability to remotely monitor the status of each Logiclave� in real time. All logged data generated by the Logiclave� controller is automatically downloaded to a database on the PC. When reviewing the data, the integrity of the data is verified before being displayedVisit the ESTS Logiclave website for more information on Logiclave Clinical Sterilizers